Display |
LCD |
|
Residual
chlorine
measurement |
Measuring range |
-0.5 to 5.5mg/l |
Resolution |
0.01mg/l |
Repeatability |
Max. 0.2% F.S |
Measurement principle |
Galvanic electrode |
Measurement objects |
Free Residual Chlrorine |
Sample flow rate |
0.5~1.0liter/min |
Input pressure |
Less than 3kg/§² |
Output pressure |
Open in air |
Sample water pH |
6~8pH(Stable) |
Sample water temperature |
0~40¡ÆC |
Sensor to Analyzer Distance |
30m(when junction box use with it) |
Analog Outputs |
Analog Outputs |
Isolated 0.00~20.00mA or 4.00~20.00mA |
Load |
MAX. 600ohm |
Communication |
Outputs |
Using isolated half duplex |
RS-485 |
Protocol |
MODBUS-RTU |
BAUD |
1200 ~ 9600 bps |
Max. |
distance Max. 1.2km |
Relay outputs |
Relay type |
SPST contact(UL) |
Switching volt. |
Max. AC250V, DC30V |
Switching Current. |
Max. 5A |
Out Relay |
3 Relay |
Function |
Alarm, Control, Timer(Performance of DIK sensor cleaning)
(Independently setup for relay output each channel are available) |
Sensor Calibration |
Method |
ZERO, SPAN |
Hold Function |
Auto Hold |
When using of sensor cleaning or sensor calibrating |
Manual Hold |
When ON of HOLD Key |
Alarm Control Relay |
Available of operating method setup when HOLD |
Analog outputs |
HOLD to previous output values |
Test Function |
Relay contact test, analog outputs test, simulation test |
|
Security Function |
Pass code |
Access to configuration settings and sensor calibration is available only by entering passcode. (Change of passcode is not available) |
Ambient Conditions |
Temperature range |
-20.0 ~ +60¡ÆC |
Storage Temp. |
-25.0 ~ +65.0¡ÆC |
Relative Humidity |
Non-condensing 10 ~ 95% |
Enclosure |
IP54(Front panel) |
Housing |
Mounting |
Panel mounting |
Dimension |
96(H) x 96(W) x 99(D)mm |
Depth |
Approximately 91mm |
Weight |
Approximately 0.5Kg |
Material |
ABS |
Power Supply |
Supply Voltage |
AC100 ~ 240V(-15% ~ +10%), 50/60Hz |
Power Consumption |
Max. 5VA |